MEDTRONIC SOFAMOR DANEK KYPHON XPANDER INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number K08A
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 06/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor applicable imaging films were returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent kyphoplasty due to primary osteoporosis and compression fracture.Intra-op, rupture of the left balloon with seven turns at 200 psi was confirmed on the screen image.When the balloon was removed and inflated outside the operation field, the rupture seemed to be a pinhole and contrast medium leakage was confirmed.No fragments of the balloon remained inside the patient's body.A new one was opened to complete the procedure.Post-op, there is no influence on the patient.
 
Additional Manufacturer Narrative
Product analysis: the complaint was indicated to be on ibt that was leaking.Both ibts were tested and it was found that one balloon had a pinhole located near the peak closest to the proximal bond.This type of pinhole puncture is consistent with the balloon coming in contact with sharp bone in the vertebral body.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
MDR Report Key7720065
Report Number1030489-2018-01050
Device Sequence Number11011857
Product Code HRX
UDI-Device Identifier00858196001154
UDI-Public00858196001154
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K041454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberK08A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2018
Initial Date Received by Manufacturer 06/25/2018
Supplement Date Received by Manufacturer08/08/2018
09/17/2018
Initial Report FDA Received Date07/25/2018
Supplement Report FDA Received Date10/03/2018
10/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-