The device history record (dhr) review confirms that the device met all material, assembly and performance specifications.During visual inspection, the delivery wire was found kinked, and the main coil was partially detached mechanically from the delivery wire indicating that the detachment was not smooth confirming the reported complaint.Functional testing could not be performed due to the condition of the returned device.It is probable that the device was damage during the clinical procedure causing the as reported events.An assignable cause of procedural factors was assigned to the as analyzed and as reported issues 'main coil failed/unable to detach' and ' main coil protrusion ' as the issue is associated with a product that meets stryker design and manufacture specification and was used in according with the direction for use (dfu), but due to the procedural and/or anatomical factors during use, the device performance was limited.It is probable that the as analyzed defect ¿coil delivery wire was kinked/bent¿ was the result of handling of the device during the preparation of the device, or during clinical procedure; therefore, an assignable cause of handling damage was assigned as the issue was due to handling of the device or portion of the device during the clinical procedure, upon removal of the device from the packaging, or preparation of the device prior to use.Expiration date - added.Device evaluated by mfg - updated to yes.Manufacturing date - added.
|