| Model Number |
METRIX AIR |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 211 and 201 mg/dl.The customer's expected fasting blood glucose test result range is 80 - 130 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.Customer also stated that he accidently left the test strip vial in the car for a whole day.The test strip lot manufacturer's expiration date is 09/13/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1 :159mg/dl date:(b)(6) 2018 time:07:36am fasting; result 2 :211mg/dl date:(b)(6) 2018 time:05:06am fasting; result 3 :201mg/dl date:(b)(6) 2018 time:05:05am fasting; result 4 :189mg/dl date:(b)(6) 2018 time:09:38pm fasting; result 5 :182mg/dl date:(b)(6) 2018 time:02:32pm fasting.
|
| |
|
Search Alerts/Recalls
|