COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 980
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Information (3190)
Date of Event 07/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, while in use on a patient, when a 980 ventilator was using the quick start ventilation mode, the patient was not ventilated because of the non-physiological default settings of the ventilator.Reportedly, the respiratory frequency, inspiratory time, and i:e (inspiratory/expiratory) ratio were not adapted to the preconfigured institutional volume settings.Reportedly, the inspiratory time was too short and the pressure alarm limit was reached immediately.Although requested, information pertaining to patient intervention and outcome (if applicable) has not been provided.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
michael collins road mervue
galway
MDR Report Key7721026
Report Number8020893-2018-00333
Device Sequence Number10951789
Product Code CBK
UDI-Device Identifier10884521171732
UDI-Public10884521171732
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K162738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number980
Device Catalogue Number980X1FRAIQC
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/13/2018
Supplement Date Received by Manufacturer08/08/2018
Initial Report FDA Received Date07/25/2018
Supplement Report FDA Received Date10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-