The customer sample was not available for evaluation.It is vital to the investigation that the customer sample be available for examination and testing.If the sample becomes available, it can be forwarded to us so further evaluation can be performed.The device history record (dhr) for the finished good was reviewed and no incident was documented that could have caused this type of non-conformance.No reported non-conformance was found from review of the records.The minimum pull test specification for this drain is 5.0 lb.Dhr demonstrated pull test results of a minimum of 12.5 lb.The non-conformance report data since july 2017 to present was reviewed and no issues that could be related to the catalog number and condition reported were found.An analysis of the complaint data was also reviewed for the same time period and demonstrates that this is the first time that this type of issue is reported from this catalog in the last 12 months.As preventive actions, the customer will be provided with a copy of the instructions for use.It is recommended to strictly follow the instructions provided in the instruction for use (ifu) data included with the product to ensure drains are properly used.According to the instructions for use (ifu).¿drains or tubing should not be handled with any instruments.This can lead to tearing, warping, or weakening and subsequent breakage of the drain.To facilitate removal of the drain, the drain and tubing portions should not be curled, pinched, over-stretched or sutured; either internally or externally.Do not suture the drain(s).Drains should be placed and removed carefully by hand only with a slow, steady pressure.Excessive force may result in breakage.During placement and removal of the drain, do not nick, cut, tear or otherwise damage the drain as this may lead to breakage.Leaving the drain implanted for any period of time which allows for tissue ingrowth around the drain and into the holes, may cause breakage on removal.¿ we will continue to monitor trends and utilize the information as part of continuous improvement.
|