| Medical Device Problem Code |
Moisture Damage (1405)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a casing/condition (moisture ingress) issue.It was alleged there was moisture behind the display.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 15-august-2018 with the following findings: investigation confirmed moisture under the lens, a leak test failed at the lens seal.Unrelated to the original complaint, it was found that the battery compartment was cracked above the grip pad with evidence of moisture in the battery compartment.
|
| |
|