JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/31/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the tecnis symfony toric multifocal intraocular lens (model zxt300 +10.5 diopter) was explanted from patient¿s left eye because of a refractive error and patient experiencing myopia.Lens model zxt225 +29.5 diopter was implanted as a replacement.There was no patient injury and no other medical or surgical intervention was required.The patient was reported doing good.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 8/1/2018.Device returned to manufacturer? yes.The product was received in a zip lock bag.Visual inspection shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional, analysis is not possible.The reported complaint cannot be confirmed.The manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr(b)(4) and capa-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7721798
Report Number9614546-2018-00720
Device Sequence Number1050014
Product Code POE
UDI-Device Identifier05050474606197
UDI-Public(01)05050474606197(17)211031
Combination Product (Y/N)Y
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/30/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model NumberZXT300
Device Catalogue NumberZXT300U105
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/29/2018
Supplement Date Received by Manufacturer08/30/2018
10/25/2020
Initial Report FDA Received Date07/26/2018
Supplement Report FDA Received Date09/20/2018
10/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Date Report Sent to FDA01/01/2005
-
-