• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 44MM/-4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 44MM/-4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6260-9-044
Medical Device Problem Codes Degraded (1153); Device Dislodged or Dislocated (2923); Material Deformation (2976)
Health Effect - Clinical Code Injury (2348)
Date of Event 06/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Additional information has been requested and if received, will be provided in the supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
 
Event or Problem Description
As reported: "you could see on x-ray where the trunnion was worn out.We took the stem out and went back with a restoration modular." review of explant photos confirm significant trunnion wear therefore confirming head-stem disassociation.
 
Additional Manufacturer Narrative
An event regarding disassociation involving a lfit v40 cocr head that was mated with an unknown stem.The event was confirmed due to the trunnion wear observed on the stem.Method & results: -device evaluation and results: visual inspection: the device was not returned but some photos of the explants were provided.Other than the biological material present on the outer surface and inside the taper, the head is unremarkable.Dimensional inspection, functional inspection & material analysis:not performed as the device was not returned.-medical records received and evaluation: no medical records were received for review with a clinical consultant -device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.The subject device has been identified to be within scope of nc and capa.Lot specific voluntary recall pfa was initiated for the lfit v40 cocr heads within scope of nc and capa pr.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analyses identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.No further investigation is required.Corrective action nc was initiated due to the occurrence rate for taper lock failure in the risk management files for specific lots of lfit v40 cocr heads.Given the potential risk of nine (9) hazards identified in the hazards and harm evaluation, voluntary product recall pfa assessment was issued.
 
Event or Problem Description
As reported: "you could see on x-ray where the trunnion was worn out.We took the stem out and went back with a restoration modular." review of explant photos confirm significant trunnion wear therefore confirming head-stem disassociation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 44MM/-4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7722202
Report Number0002249697-2018-02314
Device Sequence Number10945863
Product Code JDI
UDI-Device Identifier07613327032710
UDI-Public07613327032710
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2013
Device Catalogue Number6260-9-044
Device Lot Number6LNMKE
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/28/2018
Supplement Date Received by Manufacturer08/02/2018
Initial Report FDA Received Date07/26/2018
Supplement Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2249697-05/07/2018-003R
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-