• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100H
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/12/2018
Event Type  malfunction  
Event Description
The recipient reportedly experienced a potential device failure.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
The device passed the external visual and photographic imaging inspections.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed some of the electrical tests performed.The device passed the mechanical test performed.This is an interim report.
 
Manufacturer Narrative
The device passed the external visual and photographic imaging inspections.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.The internal visual inspection revealed that a resistor had a corroded trace.It is believed that the failure of this device was caused by a loss of hermetic seal.This is concluded from the residual gas analysis data and intrusion of the dye penetrant into the implantable cochlear stimulator (ics) inner cavity, due to a fine leak failure in the feed thru assembly.Moisture admitted into this device through this leak resulted in the corrosion of the resistive film material and the shift in resistance value of a resistor.This ultimately caused the device to cease functioning.This older device configuration is no longer manufactured.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
MDR Report Key7723015
MDR Text Key115361212
Report Number3006556115-2018-00359
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 07/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/26/2002
Device Model NumberAB-5100H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Is the Reporter a Health Professional? No
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2000
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age22 YR
-
-