When computing a correction plan with more than one correction per day and "correct axial first" selected, the hoffmann lrf software erroneously computes the axial portion of the correction, resulting in a correction plan that would be faster than intended.The surgeon overrode the (erroneous) correction time with his own calculations.We provide a surgeon override feature in case they wish to do that.
|
|
When computing a correction plan with more than one correction per day and "correct axial first" selected, the hoffmann lrf software erroneously computes the axial portion of the correction, resulting in a correction plan that would be faster than intended.The surgeon overrode the (erroneous) correction time with his own calculations.We provide a surgeon override feature in case they wish to do that.
|
|
The reported event that web application hoffmann lrf was alleged of being 'not functional' could be confirmed, since the reported outputs were replicated in software.Based on investigation, the root cause was attributed to be design related.The failure was caused by wrong calculations process.Therefore, nc and capa have been initiated to correct deviations.The nominated risks are not only valid for the hoffmann lrf hexapdod system but for all hexapod six-axis external ring fixator systems.From a clinical point of view the nominated risks are acceptable when weighted against the benefits of the devices in question.Wrong or excessive correction, correction delivered too early, correction delivered in wrong direction, or at an excessive distraction rate, the struts will need to be reoriented into a corrected position.With no web app, a correction plan will need to be created and/or applied, the frame has to be corrected, this will require adjustments to the frame.This is a normal process and usually does not cause significant harm.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|