Catalog Number 367880
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd vacutainer® lithium heparinn (lh) 158 usp units blood collection tubes exhibited erroneous results.No serious injury or medical intervention reported.
 
Additional Manufacturer Narrative
Investigation summary: bd had received samples from the customer facility for evaluation.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd had reached out to the customer and provided troubleshooting related to the issue they were observing.It was determined that the most probable scenario for their issue was that the lithium in the lithium heparin tubes was impacting the cell yield, viability, and performance due to the fact lithium can affect pbmc.The effect is more pronounced with blood aging in the tube and higher lithium concentrations.Bd has recommended that the customer use the bd cell preparation tubes (catalog 362753) for single tube separation of pbmc and for further downstream analysis.The customer acknowledged this information and the issue has been resolved.Therefore, no further evaluation of the customer samples was deemed necessary.Investigation conclusion: bd had reached out to the customer and provided troubleshooting related to the issue.The most likely root cause of the issue has been determined.At this time no further evaluation of the customer samples will be conducted.Root cause description: based on the investigation, the most likely root cause was determined to be that the lithium in the lithium heparin tubes was impacting the cell yield, viability, and performance due to the fact lithium can affect pbmc.
 
Event or Problem Description
It was reported that bd vacutainer® lithium heparin (lh) 158 usp units blood collection tubes exhibited erroneous results.No serious injury or medical intervention reported.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd had reached out to the customer and provided troubleshooting related to the issue they were observing.It was determined that the most probable scenario for their issue was that the lithium in the lithium heparin tubes was impacting the cell yield, viability, and performance due to the fact lithium can affect pbmc.The effect is more pronounced with blood aging in the tube and higher lithium concentrations.Bd has recommended that the customer use the bd cell preparation tubes (catalog 362753) for single tube separation of pbmc and for further downstream analysis.The customer acknowledged this information and the issue has been resolved.As no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.As there was no sample or photo available for evaluation, a root cause could not be determined.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that bd vacutainer® lithium heparinn (lh) 158 usp units blood collection tubes exhibited erroneous results.No serious injury or medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® LITHIUM HEPARINN BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key7723720
Report Number1917413-2018-03115
Device Sequence Number16541310
Product Code JKA
UDI-Device Identifier50382903678803
UDI-Public50382903678803
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K901449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/31/2019
Device Catalogue Number367880
Device Lot Number7340932
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2018
Initial Date Received by Manufacturer 07/05/2018
Supplement Date Received by Manufacturer07/05/2018
07/05/2018
Initial Report FDA Received Date07/26/2018
Supplement Report FDA Received Date12/26/2018
12/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-