• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0035431030
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.During the analysis of the returned device, it was revealed that the main coil was broken and stretched.In addition, the coil delivery wire was kinked due to handling.During the functioning test, delivery wire moved independently to the main coil and the main coil was unable to be removed from the introducer sheath.Additional information provided by the customer indicated that no anomalies were noted to the device prior to use and there is no indication of a use error.It is probable that the device was damaged during the clinical procedure causing the reported event.An assignable cause of procedural factors will be assigned to the reported event and to the analyzed main coil broken and stretched, and coil introducer sheath friction.As the issue is associated with a product that meets stryker design and manufacture specifications and was used according to the dfu (direction for use) but due to procedural and/or anatomical factors during use, the device performance was limited.
 
Event or Problem Description
During the analysis of the returned device, it was revealed that the main coil was broken.No clinical consequences reported to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET HELICAL NANO 1MM X 3CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7724850
Report Number3008881809-2018-00324
Device Sequence Number1041040
Product Code HCG
UDI-Device Identifier04546540698018
UDI-Public04546540698018
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Catalogue NumberM0035431030
Device Lot Number20193652
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2018
Initial Report FDA Received Date07/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-