• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION ZOLL IVTM QUATTRO CATHETER; CENTRAL VENOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION ZOLL IVTM QUATTRO CATHETER; CENTRAL VENOUS CATHETER Back to Search Results
Model Number IC-4593
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The zoll catheter was returned to zoll on 03 jul 2018 for investigation.However, investigation is still in progress.A supplemental report will be filed when investigation has been completed.
 
Event or Problem Description
Approximated 26 hours during ivtm therapy, the thermogard console displayed an airtrap alarm and the user observed the saline bag was low.Upon further inspection, the user observed fluid underneath the console.Following this, the user discontinued ivtm therapy.The user was able to clear the airtrap alarm.Adjunct therapy was administered.No known impact or patient consequence was reported.
 
Additional Manufacturer Narrative
The reported catheter leak was not confirmed.Visual inspection of the returned catheter was performed.The balloons were examined and there were no tears, balloon bond detachment or rupture noted.Functional testing of the returned quattro catheter (lot no.79089) was performed.All infusion ports and extension tubes were flushed without resistance.The catheter was connected to a pressurized inflation device, the balloons fully inflated when pressurized up to 100 psi without any issues.The balloons did not leak during inflation and deflation.No leak was observed.All lumens flushed as intended.During manufacturing, all catheters are 100 % inspected.Only units that pass are moved to the next process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZOLL IVTM QUATTRO CATHETER
Common Device Name
CENTRAL VENOUS CATHETER
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
MDR Report Key7725211
Report Number3010617000-2018-00753
Device Sequence Number1118418
Product Code NCX
UDI-Device Identifier00849111075138
UDI-Public00849111075138
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K101987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIC-4593
Device Catalogue Number8700-0660-01
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/03/2018
Initial Date Received by Manufacturer 06/27/2018
Supplement Date Received by Manufacturer08/17/2018
Initial Report FDA Received Date07/26/2018
Supplement Report FDA Received Date08/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age35 YR
Patient Weight82
-
-