• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAL
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hyperglycemia (1905); Skin Irritation (2076); Diabetic Ketoacidosis (2364)
Event Date 06/24/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call that they hospitalized as a result of high blood glucose, infection on site and diabetes ketoacidosis on (b)(6) 2018 with the blood glucose of 500 mg/dl.The customer treated with the antibiotics and ivy.Troubleshooting was done for site issues and declined for insulin pump.The insulin pump will not be returned for analysis.
 
Manufacturer Narrative
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accuracy test at 0.08750 inches.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.Device uploaded properly using carelink.Device had minor scratched display window, cracked display window, cracked case at display window corner, cracked battery tube threads, cracked reservoir tube lip and a missing end cap sticker.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAL
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key7725383
MDR Text Key115188032
Report Number3004209178-2018-96277
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169404328
UDI-Public(01)00643169404328
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751NAL
Device Catalogue NumberMMT-751NAL
Device Lot NumberA2751NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received12/02/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age29 YR
Patient Weight160
-
-