MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 133MV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 133MV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133MV-14-T
Medical Device Problem Codes Fluid/Blood Leak (1250); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Device labeling addresses: patients should be advised that the tissue expanders may deflate, and require replacement surgery.Deflation occurs when saline leaks through a damaged injection site or a damaged tissue expander envelope.
 
Event or Problem Description
Healthcare professional reported a left side deflation.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: opening curve, yellow particles, and yellow biological tissue.Leak test was performed and found opening on anterior.A microscopic analysis was performed which identified sharp edge opening assessed as unidentified tear opening, opening striated in the anterior side, also a striated break in the suture tabs and missing piece of the suture tab.A dimension measurement of the shell was performed which identified the thickness within specification.The fill test inspection was performed, with the result of no blockage.Based on the device analysis the final assessment is: a sharp opening in the anterior side due to an unidentified (tear) opening, a striated edge opening on anterior side due to surgical damage, a striated break in the suture tab, and a missing suture tab due to inconclusive.Device labeling: natrelle® 133 tissue expanders are temporary devices, and are not to be used for permanent implantation or beyond 6 months.Tissue expanders should be removed once adequate tissue has developed.Tissue expansion in breast reconstruction typically requires four to six months.The total expansion period will vary depending on patient tolerance, tissue tolerance, and desired tissue expansion.
 
Event or Problem Description
Device was implanted for more than 6 months.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133MV TISSUE EXPANDER WITH SUTURE TAB
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key7725692
Report Number9617229-2018-05475
Device Sequence Number11080091
Product Code LCJ
UDI-Device Identifier10888628004436
UDI-Public10888628004436
Combination Product (Y/N)N
Initial Reporter StateWA
PMA/510(K) Number
K102806
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/21/2021
Device Catalogue Number133MV-14-T
Device Lot Number3044461
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2018
Initial Date Received by Manufacturer 06/28/2018
Supplement Date Received by Manufacturer08/27/2018
Initial Report FDA Received Date07/27/2018
Supplement Report FDA Received Date09/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-