DEPUY MITEK MILAGRO INTERFERENCE SCREW, 9 X 30MM; ORTHOPAEDIC BONE SCREW, BIOABSORBABLE
|
Back to Search Results |
|
| Catalog Number |
231825 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/01/2014
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint devices were received and evaluated.Visual observation reveals that the anchor is broken in the middle, confirming the reported complaint.No damage to the threads of the proximal broken end.The distal end was received in multiple fragments.The root cause of this failure cannot be determined.A batch record review has been conducted for this batch to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any type for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
| |
|
Event or Problem Description
|
|
It was reported by the affiliate in vietnam that during a kidney surgery, it was observed that the milagro interference screw, 9 x 30mm was broken when the surgeon screwed it into the tibial tunnel.The surgeon had to use another screw to keep the gaft stability into tunnel.It was not reported if there were any delays in the procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|