• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND INGENIO; IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND INGENIO; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number K172
Medical Device Problem Codes High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877); Capturing Problem (2891)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event or Problem Description
Boston scientific received information that this pacemaker exhibited spikes in impedances on the right ventricular (rv) lead.One value was noted to be out of range at greater than 2000 ohms.Further review showed that there was noise on the rv channel as well which was oversensed.The noise was due to the minute ventilation signal, so boston scientific technical services recommended programming the feature off.Variable thresholds were also observed.The patient is not dependent.The system remains in service and the patient is being monitored.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENIO
Common Device Name
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
MDR Report Key7725867
Report Number2124215-2018-13032
Device Sequence Number11125888
Product Code LWP
UDI-Device Identifier00802526516818
UDI-Public(01)00802526516818(17)20160417
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/17/2016
Device Model NumberK172
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2018
Initial Report FDA Received Date07/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4137; K172
Patient Age71 YR
-
-