• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OBTURATOR,INTEL 4MM A'SP FOR; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OBTURATOR,INTEL 4MM A'SP FOR; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 4356
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
Date of Event 05/19/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there were black debris in the joint.Doctors thinks there is shredding of the ring.All of the material was flushed out of the patient.No patient injury reported.
 
Additional Manufacturer Narrative
One 4mm obturator was returned for evaluation.Visual assessment of the device showed no defects that would contribute to the reported complaint.The device was tested with a 6mm high flow double valve cannula and was found to function as intended.This investigation could not identify any evidence of product contribution to the reported complaint.Has been corrected for the initial mdr, the date of the inital mdr should be read as (b)(4) 2018 instead of (b)(4) 2018.
 
Event or Problem Description
It was reported that there were black debris in the joint.Doctors thinks there is shredding of the ring.All of the material was flushed out of the patient.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OBTURATOR,INTEL 4MM A'SP FOR
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key7726236
Report Number1219602-2018-00938
Device Sequence Number19173806
Product Code NBH
UDI-Device Identifier03596010129024
UDI-Public(01)03596010129024(10)50697610
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2018
1 Device was Involved in the Event
2 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4356
Device Catalogue Number4356
Device Lot Number50697610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2018
Supplement Date Received by Manufacturer06/19/2018
Initial Report FDA Received Date07/27/2018
Supplement Report FDA Received Date08/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number0
-
-