| Model Number |
4356 |
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
05/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there were black debris in the joint.Doctors thinks there is shredding of the ring.All of the material was flushed out of the patient.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One 4mm obturator was returned for evaluation.Visual assessment of the device showed no defects that would contribute to the reported complaint.The device was tested with a 6mm high flow double valve cannula and was found to function as intended.This investigation could not identify any evidence of product contribution to the reported complaint.Has been corrected for the initial mdr, the date of the inital mdr should be read as (b)(4) 2018 instead of (b)(4) 2018.
|
| |
|
Event or Problem Description
|
|
It was reported that there were black debris in the joint.Doctors thinks there is shredding of the ring.All of the material was flushed out of the patient.No patient injury reported.
|
| |
|