|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint devices were received and evaluated.Visual observation for both devices confirms the distal tip of the anchors are broken, but are held in place by the suture.The anchors appear to be used and there are evidence of tissue debris found on the anchors and the tip of the shaft.This type of anchor breakage at its distal tip is consistent with historical investigations in which it was determined that the root cause was likely a combination of torque and bending, a user technique issue.Other than this possibility, we cannot discern a root cause for this failure.A batch review was conducted to determine if there were any internal processing issues, which would have contributed to the nature of the product complaint.Our results indicate that this batch was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the mitek complaints system revealed no other complaints of any type for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|