| Model Number |
TN,DXH 800 HEMATOLOGY SYSTEM |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Test Result (2695)
|
| Date of Event |
09/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer technical specialist (cts) was on the phone with the customer on (b)(4) 2017.The cts guided the customer through cleaning the aspiration probe, zapping the apertures and priming the instrument.The customer then ran background checks which passed.The cause of the reported issue could not be identified.Beckman coulter is conducting a retrospective review of its global service records, to ensure that the records had been adequately evaluated for potential complaint information and any warranted follow-up action, including mdr reporting. this mdr report is a result of that retrospective review. an improvement for plt flagging was developed and released as a part of a software update to the dxh800 that is designated v3.2.1 on 22-dec-2017; approximately 11.5% of the worldwide product installed base has been upgraded as of 13-jul-2018.Due to an increase in the number of complaints received related to unflagged erroneously high plt patient results released from the lab, bec anticipates that a field action will be warranted.The anticipated field action will include a software patch for the improvement for the plt flagging component of v3.2.1, which will expedite installation of this improvement for the rest of the installed base.Section a patient information was not provided by the customer.Bec internal identifier (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer stated the platelet (plt) results were not matching on reruns.There was no report of change or effect to patient treatment in connection with the reported event.Patient data was not provided by the customer for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up 01: an improvement for plt flagging was developed and released as a part of a software update to the dxh800 that is designated v3.2.1 on 22-dec-2017; approximately 11.5% of the worldwide product installed base has been upgraded as of 13-jul-2018.Due to an increase in the number of complaints received related to unflagged erroneously high plt patient results released from the lab, bec has determined that a field action is warranted.Field action (fa-33718) was approved on 30-jul-2018.The field action includes a software patch for the improvement for the plt flagging component of v3.2.1, which will expedite installation of this improvement for rest of the installed base.Section h2 - additional information has been selected.Section h7 - recall, notification and modification have been selected.Section h9 - number has been updated to reflect the respective information for this event.Additional information: the recall (fa-33718) includes notification to the customer and correction of the issue via software update.Bec internal identifier: (b)(4).
|
| |
|
Search Alerts/Recalls
|