| Model Number |
TN, LH780 ANALYTICAL STATION |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Test Result (2695)
|
| Date of Event |
03/16/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On 25-mar-2016 the field service engineer (fse) was at the customer site and observed that the reported issue was occurring in manual mode.The fse took apart and cleaned the bsv (blood sampling valve) to resolve the issue.The repairs were verified per established procedures.The cause of the high plt results is attributed to an obstructed bsv.Section a - patient information was not provided by the customer.Beckman coulter is conducting a retrospective review of its global service records, to ensure that the records had been adequately evaluated for potential complaint information and any warranted follow-up action, including mdr reporting. this mdr report is a result of that retrospective review. (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported that their coulter lh 780 hematology analyzer was generating erratic platelet (plt) results and the controls were also biased high for platelets.The customer had not reported any erroneous results outside of the laboratory.There was no report of change or effect to patient treatment in connection with the reported event.The issue was reported as an ongoing issue for a two week period by the customer but they had not called to report it previous to this call.
|
| |
|