| Model Number |
TN, LH780 ANALYTICAL STATION |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Test Result (2695)
|
| Date of Event |
01/29/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) was at the customer site on (b)(6) 2016 and observed that solenoid sl10 (white blood cell count) was not firing.The solenoid was replaced to resolve the reported issue.The cause of the high plt results is attributed to a defective solenoid sl10.Patient information not provided by the customer.Beckman coulter is conducting a retrospective review of its global service records, to ensure that the records had been adequately evaluated for potential complaint information and any warranted follow-up action, including mdr reporting. this mdr report is a result of that retrospective review. bec internal identifier - case-(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported that the rbc (red blood cell) and plt (platelet) results generated on their coulter lh 780 hematology analyzer are erratic.The results were described to have varied from zero to very high.There was no report of change or effect to patient treatment in connection with the reported event.Patient data was not provided by the customer for this event.
|
| |
|
Search Alerts/Recalls
|