• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER AC·T DIFF 2 ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER AC·T DIFF 2 ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number AC*T DIFF 2
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Test Result (2695)
Date of Event 11/08/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On (b)(6) 2016, the field service engineer (fse) was at the customer site and observed an abnormal plt histogram and found that plt counts were elevated on the ac*t diff 2 instrument.The fse found that the probe was restricted and replaced the probe to resolve the issue.The repairs were verified by the fse.The failure mode is attributed to the probe which was replaced by the fse.Patient information was not provided by the customer.Beckman coulter is conducting a retrospective review of its global service records, to ensure that the records had been adequately evaluated for potential complaint information and any warranted follow-up action, including mdr reporting.  this mdr report is a result of that retrospective review.  (b)(4).
 
Event or Problem Description
The customer reported that their coulter ac·t diff 2 analyzer generated erroneous high platelet (plt) results on three patient samples.The customer also stated that controls were failing for plts after the event was identified.There was no report of erroneous results reported out of the lab from the provided information.There was no report of change or effect to patient treatment in connection with the reported event.Patient data was not provided by the customer for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER AC·T DIFF 2 ANALYZER
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd
brea CA 92821
MDR Report Key7726894
Report Number1061932-2018-01018
Device Sequence Number19058427
Product Code GKZ
UDI-Device Identifier15099590264796
UDI-Public(01)15099590264796(11)NO-DATA
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K990352
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAC*T DIFF 2
Device Catalogue Number6605500
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/08/2016
Initial Report FDA Received Date07/27/2018
Patient Sequence Number1
-
-