MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number TN, LH780 ANALYTICAL STATION
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Test Result (2695)
Date of Event 11/09/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On 9 nov 2016, the field service engineer (fse) was at the customer site and inspected the lh780 instrument but was not able to find an instrument malfunction.The fse replaced tubing and cleaned the instrument as preventive maintenance.The fse verified that the instrument was working per specifications.The failure mode is unknown.Beckman coulter is conducting a retrospective review of its global service records, to ensure that the records had been adequately evaluated for potential complaint information and any warranted follow-up action, including mdr reporting.  this mdr report is a result of that retrospective review.  (b)(4).
 
Event or Problem Description
The customer reported that their coulter lh 780 hematology analyzer is generating high platelet (plt) results.There was no report of change or effect to patient treatment in connection with the reported event.Patient data was not provided by the customer for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 780 HEMATOLOGY ANALYZER
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd
brea CA 92821
MDR Report Key7726898
Report Number1061932-2018-01015
Device Sequence Number16397630
Product Code GKZ
UDI-Device Identifier15099590282431
UDI-Public(01)15099590282431(11)NO-DATA
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K061616
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTN, LH780 ANALYTICAL STATION
Device Catalogue Number723585
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/09/2016
Initial Report FDA Received Date07/27/2018
Patient Sequence Number1
-
-