| Model Number |
TN, LH 500 (108-132/48-62) |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Test Result (2695)
|
| Date of Event |
01/30/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On 30 jan 2017, the field service engineer (fse) was at the customer site and replaced the rbc bath and aperture on the coulter lh 500 hematology analyzer to resolve the issue.The repairs were verified by the fse.The failure mode is the rbc bath and aperture that were replaced by the fse.Patient information was not provided by the customer for this event.Beckman coulter is conducting a retrospective review of its global service records, to ensure that the records had been adequately evaluated for potential complaint information and any warranted follow-up action, including mdr reporting. this mdr report is a result of that retrospective review. bec internal identifier - (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported that the coulter lh 500 hematology analyzer generated erroneous high platelet (plt) results on patient samples.There was no report of erroneous results reported out of the lab from the provided information.There was no report of change or effect to patient treatment in connection with the reported event.Instrument printouts of patient results or controls were not provided for this event.
|
| |
|