| Model Number |
CC4204A +18.5 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.Device evaluation: lens was returned in liquid, in lens case/vial.Visual inspection found a torn optic and haptic.Work order search: no similar complaint type events were reported for units within the same lot.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a cc4204a, +18.5 diopter, implantable collamer lens tore during the loading process.There was no patient contact and the backup lens was implanted.Cause of event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: (b)(4).
|
| |
|
Search Alerts/Recalls
|