U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SUBQ; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SUBQ; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 6996SQ58
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Date of Event 07/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: h601h27 heart ring, implanted: (b)(6) 1996; ddbc3d1 icd, implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead alerted for low impedance on the subcutaneous high voltage coil.The lead was revised and removed.A new intravenous lead was implanted.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the right ventricular defibrillation coil was below the expected lower range.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUBQ
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7727759
Report Number2649622-2018-12303
Device Sequence Number904604
Product Code LWS
UDI-Device Identifier00613994568021
UDI-Public00613994568021
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/06/2018
Device Model Number6996SQ58
Device Catalogue Number6996SQ58
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/04/2018
Supplement Date Received by Manufacturer07/30/2018
08/08/2018
08/08/2018
Initial Report FDA Received Date07/27/2018
Supplement Report FDA Received Date08/02/2018
10/03/2018
10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4968-35 X2 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age25 YR
-
-