| Model Number |
72201389 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that sales representative noticed that there is a friction point and drill bit will not slide through drill guide.No back up device was available.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Examination was not possible, as the device has not been returned.The investigation could not draw any conclusions about the reported event without the return of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received by the manufacturer has identified that this event should be re-evaluated for mdr reporting: the event was reported because there was no back up device available during the procedure, however per communication with the sales representative, it was confirmed that there was a smith and nephew back up device available to complete the procedure succesfully therefore the reassessment determined that the issue does not meet the threshold for reporting and is a non-reportable event.This report was made based upon information which smith & nephew had not been able to investigate or verify prior to the required reporting date.
|
| |
|