| Model Number |
S-LLF200TG |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the returned device revealed that the fiber was returned in two pieces.The fiber distal tip is fractured.The fiber is broken within the fiber blue outer sheath, and detached at 9.5 inches from distal tip.The length of the second piece including the connector is 10 feet and 9.5 inches.The fiber exhibits signs of melting at the location of fracture for first piece.The fiber was tested with hene laser fixture, and the aim beam is visible at the break.Based on the device analysis, the probable cause of the failure is handling damage.An evaluation conclusion code of cause traced to user was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
Analysis of the return device revealed that the fiber is broken within the fiber blue outer sheath, a fracture on the distal tip of the fiber, and a detachment of the fiber 9.5 inches from the distal tip.There were no patient complications reported due to this event.
|
| |
|