AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) SUREFLEX FIBERS; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number S-LLF200TG
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis of the returned device revealed that the fiber was returned in two pieces.The fiber distal tip is fractured.The fiber is broken within the fiber blue outer sheath, and detached at 9.5 inches from distal tip.The length of the second piece including the connector is 10 feet and 9.5 inches.The fiber exhibits signs of melting at the location of fracture for first piece.The fiber was tested with hene laser fixture, and the aim beam is visible at the break.Based on the device analysis, the probable cause of the failure is handling damage.An evaluation conclusion code of cause traced to user was assigned to this investigation.
 
Event or Problem Description
Analysis of the return device revealed that the fiber is broken within the fiber blue outer sheath, a fracture on the distal tip of the fiber, and a detachment of the fiber 9.5 inches from the distal tip.There were no patient complications reported due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFLEX FIBERS
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
MDR Report Key7728405
Report Number2937094-2018-00811
Device Sequence Number19183702
Product Code GEX
UDI-Device Identifier00878953004822
UDI-Public00878953004822
Combination Product (Y/N)N
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/09/2020
Device Model NumberS-LLF200TG
Device Catalogue NumberS-LLF200TG
Device Lot Number704B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2018
Initial Report FDA Received Date07/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-