• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number E2781R-28ASP
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during procedure, the electrode worked fine in the beginning but then suddenly the electrode tip broke and piece fell into patient's cavity.The piece was removed from the cavity.New device was used to complete the procedure.No patient harm.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned product did not meet specification as received.Visual inspection found the tip of the electrode to have broken off.The tip fractured.The investigation found the tip to have broken off.The broken tip was returned.The broken tip was inspected under magnification.The broken off section had stress fractures.The investigation identified the root cause of the reported event to be the customer applying excessive force to the electrode tip.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7728417
Report Number1717344-2018-00991
Device Sequence Number806016
Product Code GEI
UDI-Device Identifier10884524000701
UDI-Public10884524000701
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K944244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2022
Device Model NumberE2781R-28ASP
Device Catalogue NumberE2781R-28ASP
Device Lot Number73050116X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Initial Date Received by Manufacturer 06/27/2018
Supplement Date Received by Manufacturer08/08/2018
09/18/2018
Initial Report FDA Received Date07/27/2018
Supplement Report FDA Received Date10/03/2018
10/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-