| Model Number |
1885076HSE |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation was performed as the device has not been returned at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported via a manufacturer representative that the bur wasn't working well and the tool broke at the end when trying to detach it from the microdebrider.The tool was replaced and the microdebrider was changed.It was noted that patient symptoms were unknown and there was no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the returned device indicated that the inner assembly would spin freely by hand.When viewed under magnification, there was damage to the hubs that is consistent with improper loading: dimples on the front hub prior to the locking area caused by the handpiece locking mechanism; locking area deformation caused by the back side of the front collet of the handpiece; and deformation of the inner hub chevrons caused by the handpiece drive mechanism.There were no loose components.Even with the deformation of the hubs, functionally, the bur loaded securely into a handpiece, ran at 12,000 rpm in forward mode, and cut saw bone with no issues.There was no breakage of the bur.There was an out of specification condition identified as it relates to the complaint (due to physical damage).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|