No patient involvement was reported.Lot number provided, 8097, is supplier lot.Synthes lot number is 4727743.Device is an instrument and is not implanted/explanted.Initial reporter is synthes sales consultant.A review of the device history records has been requested.Subject device has been received and is currently in the evaluation process.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that during a field inventory inspection on july 2, 2018, a self-retaining screwdriver for locking screw was found to have a broken tip.There was no patient involvement.This report is for one (1) self-retaining screwdriver for locking screw this is report 1 of 1 for (b)(4).
|
|
Device history records review was completed for part# 387.284, lot# 4727743.Supplier: (b)(4).Release to warehouse date: feb 24, 2004.No non-conformance reports were generated during production.The material was reviewed, and the hardness value was confirmed to meet the specification with no (relevant) non-conformance noted.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.: product development investigation was completed.The screwdriver (part # 387.284, lot # 4727743, mfg 24-feb-2004) was received at us customer quality with two of the sharp edges of the tip broken off.There are no other visual defects on the screwdriver.This is consistent with the reported complaint condition, thus confirming the complaint.Dimensional analysis was completed, the outer diameter the shaft distal from the tip measured 4.48 mm.This is within specification of 4.46 to 4.50 mm, based on the relevant drawing.The outer diameter of the shaft at the distal tip close to the fracture point measured 3.35 mm.This is within specification of 3.30 to 3.38 mm, based on the relevant drawing.The complaint condition is confirmed as the screwdriver (part # 387.284, lot # 4727743) was received with two of the sharp edges of the tip broken off.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces (such as being dropped, struck off axis or damaged during sterile processing).No new malfunctions were observed during the course of this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Corrected data: lot#, serial #.Mfr site: manufacturing facility and address.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|