Catalog Number 7770728
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that intra-operatively 2 cages were broke while impacted.The peek part was cracked off where the implant inserter connected.The device came in contact with the patient.No fragment of broken cage were remained in patient.No patient complications were reported due to this event.
 
Additional Manufacturer Narrative
Product analysis: visual review of the implant confirmed the ¿ears of the -03 top component deformed and broken.Microscopic examination of the -03 peek top component scalloped features did identify and witness marks, suggesting possible physical obstruction during attempted implantation which prevented full insertion into the vertebral disc space.The above observations are consistent with implant fracture due to brittle overload during attempted implantation.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATE SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7729644
Report Number1030489-2018-01068
Device Sequence Number19478040
Product Code MAX
UDI-Device Identifier00643169430655
UDI-Public00643169430655
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K172199
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/26/2024
Device Catalogue Number7770728
Device Lot Number0516248W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Initial Date Received by Manufacturer 06/28/2018
Supplement Date Received by Manufacturer08/08/2018
09/14/2018
Initial Report FDA Received Date07/28/2018
Supplement Report FDA Received Date10/03/2018
10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
Patient Weight97
-
-