Device Problem Failure to Align (2522)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/12/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluation by manufacturer: field service engineering visited the customer to address to address the reported event.Fse was able to confirm the error in the error log.During further troubleshooting, fse determined that the error message was caused by needle being out of alignment in the x-axis with a sample loader.Fse then cleaned and lubricated both rails in the sample loader and cleaned the sampling loader.Fse then cleaned and lubricated the sampler z and y axis.Fse ran 10 patients to verify proper alignment and chromatograms.Fse also calibrated and ran quality controls with acceptable results.The instrument was verified as operational.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018.There was one similar complaint identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: the 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.The 710 z1-axis error: an abnormality occurred in the up and down movement of the sampling needle.If this occurs during a stat assay, check that the container setting (cup or tube) is correctly set.The error also occurs when the sample vial was not recognized as a primary tube, due to the disoriented sample sensor.The most probable cause of the reported event was the sample needle was out of alignment.
 
Event Description
A customer reported getting 710 z1-axis error messages on the g8 instrument.The customer reported that the instrument was down.Field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7730043
MDR Text Key115903737
Report Number8031673-2018-00727
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 07/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/28/2018
Distributor Facility Aware Date07/12/2018
Device Age6 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer07/28/2018
Date Manufacturer Received07/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-