| Model Number |
MMT-523RNAH |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hyperglycemia (1905); Hypoglycemia (1912)
|
| Date of Event |
01/13/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer called to report high blood glucose of 401 mg/dl.Troubleshooting determined the insulin pump is working as designed.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test, basic occlusion test, occlusion test, prime test, rewind test and excessive no delivery test.
|
| |
|