ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of submission: 29-jul-2018.Device evaluation: the device has been returned and evaluated by product analysis on 24-jul-2018 with the following findings: during the investigation, the keypad was found to be intact and all keys were found to be unresponsive.The pumps cover was removed, and internal moisture was found on the printed circuit board and the keypad connector.The keypad cover was removed, and no contamination was found under the key contacts.Unrelated to the original complaint moisture was found behind the display lens.A leak test failed due to a base crack on the front of pump between the case seal and the keypad.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a button/keypad (tactile changes/unresponsive) issue.The reporter alleged that the up, down, and ok/enter button's were under responsive on the keypad.There was no indication that the product caused or contributed to an adverse event.The complaint is being reported because the issue has the ability to result in inadvertent or incorrect insulin delivery or the inability to use the pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7730418
Report Number2531779-2018-13700
Device Sequence Number2542512
Product Code OYC
UDI-Device Identifier10840406100372
UDI-Public10840406100372
Combination Product (Y/N)Y
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2018
Was the Report Sent to FDA? Yes
Device Age22 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2018
Initial Report FDA Received Date07/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age12 YR
-
-