MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL PERCEPTA QUAD CRTP PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL PERCEPTA QUAD CRTP PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION Back to Search Results
Model Number W4TR01
Device Problems Nonstandard Device (1420); Loss of Power (1475); Reset Problem (3019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/05/2018
Event Type  Injury  
Manufacturer Narrative
This device was reported as included in the field action. Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action. It is included in the field action in the abundance of caution. If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the cardiac resynchronization therapy pacemaker (crt-p) exhibited a power on reset (por). A device update, delivered via in-clinic interrogation, was performed. The crt-p remains in use. No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePERCEPTA QUAD CRTP
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7731003
MDR Text Key115364567
Report Number9614453-2018-02677
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Remedial Action Recall
Type of Report Initial,Followup
Report Date 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/14/2019
Device Model NumberW4TR01
Device Catalogue NumberW4TR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured10/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2413-2018

Patient Treatment Data
Date Received: 07/30/2018 Patient Sequence Number: 1
Treatment
5076-52 LEAD, 5076-58 LEAD, 429888 LEAD
-
-