Catalog Number 197618140
Medical Device Problem Codes Loose or Intermittent Connection (1371); Migration (4003)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); Not Applicable (3189)
Date of Event 07/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the reclaim stem appeared to be drastically undersized and it rotated into a highly anteverted position within the proximal femur.It was also stated that the stem was loose at bone to implant interface.Doi: unknown; dor: (b)(6) 2018; left hip.
 
Additional Manufacturer Narrative
(b)(4).No device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM DISTAL TAPERED 18X140
Common Device Name
RECLAIM IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
MDR Report Key7731051
Report Number1818910-2018-65405
Device Sequence Number1405637
Product Code KWA
UDI-Device Identifier10603295077077
UDI-Public10603295077077
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K102080
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number197618140
Device Lot Number665255
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/03/2018
Supplement Date Received by Manufacturer08/27/2018
Initial Report FDA Received Date07/30/2018
Supplement Report FDA Received Date08/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
-
-