Model Number 179078P
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Health professional.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, they noticed a chip at the bottom of the trocar.The camera was used to try and visualize the piece to see if it was broken off in the body.When the trocar was carefully removed, the broken piece was seen in the subcutaneous tissue in the trocar site.The piece was then removed and all parts of the trocar were collected and quarantined.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Duplicate of (b)(4).
 
Additional Manufacturer Narrative
This report 9612501-2018-01373 was sent in error as a duplicate for mfr report number: 9612501-2018-01343, any additional information received in the future will be captured and submitted under the report number 9612501-2018-01343¿.Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned products noted that the distal end of the cannula was cracked, and a piece was disengaged.The disengaged piece was received in a sample jar.The trocar, circular seal, envelope seal and bladed obturator all appeared intact.Functional evaluation is not available.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there was no assembly component related condition.Replication of the disengaged cannula tip may occur if the device is mishandled or handled roughly during application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPORT PLUS
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7731625
Report Number9612501-2018-01373
Device Sequence Number16837045
Product Code GCJ
UDI-Device Identifier20884521082660
UDI-Public20884521082660
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K954108
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Model Number179078P
Device Catalogue Number179078P
Device Lot NumberJ8A1228X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2018
Initial Date Received by Manufacturer 07/13/2018
Supplement Date Received by Manufacturer07/27/2018
08/08/2018
09/13/2018
Initial Report FDA Received Date07/30/2018
Supplement Report FDA Received Date08/13/2018
10/02/2018
10/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-