| Model Number |
MZ9271 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: pri tubing; 10ml bd syringe ref (b)(4), lot number 724212a, exp 2020-08-20, 0.9% nacl, therapy date (b)(6) 2018.The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
| |
|
Event or Problem Description
|
|
The customer reported the pump module alarmed for high pressure during an epinephrine drip.The patient¿s blood pressure was very labile and a second tubing and pump module was prepared to replace the infusion that alarmed.When changing the epinephrine line, the connection of the set to the picc was wet from a fluid leak; therefore, the new epinephrine infusion was connected to another port on the picc.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided.The customer¿s report of a leak at the connection was not confirmed.Visual inspection of the set did not reveal any discernible signs of damage or abnormalities.Functional testing resulted in no leaking.A test pump infusion reached completion without any pump alarms or irregularities noted.The root cause of the reported failure was not determined.
|
| |
|
Event or Problem Description
|
|
The customer reported the pump module alarmed for high pressure during an epinephrine drip.The patient¿s blood pressure was very labile and a second tubing and pump module were prepared to replace the infusion that alarmed.When changing the epinephrine line, the connection of the set to the picc was wet from a fluid leak; therefore, the new epinephrine infusion was connected to another port on the picc.There was no report of patient harm.
|
| |
|