| Model Number |
H965430500 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).A fully constrained wallstent biliary uncovered stent and delivery system were received for analysis.Visual examination of the returned device found the stent and tip were pulled into the outer sheath.The stainless steel tube was kinked.Functional evaluation revealed that it was not possible to deploy the stent using the delivery system as received; however, the stent was deployed by cutting the outer sheath and pulling the inner member distally.Visual examination of the stent revealed that there were uncrossed and bent stent wires at one end.There were no other issues with the stent noted and no other issues with the delivery system were noted.The damages noted were consistent with the excessive force applied to the device.Taking all available information into consideration, the investigation concluded that the observed failures and the reported event are likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device.Therefore, the most probable root cause is operational context.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.A search of the complaint database confirmed that no other similar complaints exist for the specified lot.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a wallstent biliary uncovered stent was to be used to treat a 3cm malignant stenosis in the lower part of the bile duct during a stent placement procedure performed on (b)(6) 2018.According to the complainant, during procedure, the stent failed to deploy.Reportedly, several attempts were made to move the handle of the delivery system to start the deployment, but was unsuccessful.The stent was fully covered with the outer sheath when removed from the patient.The procedure was complete with another wallstent biliary stent.There were no patient complications reported as a result of this event.The patient¿s condition at the conclusion of the procedure was reported to be stable.Note: this event has been deemed an mdr-reportable event based on investigation results which revealed uncrossed and bent stent wires at one end.
|
| |
|
Search Alerts/Recalls
|