| Model Number |
VP-708-X |
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open bypass operation, while they were suturing the patch with a continuous sewing technique on the vessel, the middle part of the needle broke.The surgeon used another suture to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of four devices.Visual inspection and functional testing confirmed there were no abnormalities that would have caused or contributed to the reported condition.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.A definitive root cause could not be determined with regard to the reported condition.However, probable causes could be attributed to mishandling the needle by grasping it at the tip or swaged end.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|