(b)(4).Date sent: 08/10/2018.The analysis results found that the lx107 device was received with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon functional testing of the device, the instrument loaded, retained and deployed 6 clips as intended.The instrument was fully functional and conforming to our manufacturing requirements.The event described could not be confirmed as the device performed without any difficulties noted.The batch history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment. the certificate records are accessible through external manufacturing. .
|