• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR(R) PLUS DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR(R) PLUS DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number CRPLUS
Medical Device Problem Code Failure to Analyze Signal (1539)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/10/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio determined that the cause of the reported issue was process residue on a capacitor, designator c157, located on the analog pcb assembly.This resulted in an improper voltage at integrated circuit (ic) designator u46 on leg 7.The customer was provided with a replacement device.
 
Event or Problem Description
The customer contacted physio-control to report that their device had a service wrench icon on its readiness display.There was no patient use associated with the reported event.Upon evaluation of the customer¿s device, physio-control found that the device does not recognize a shockable rhythm and does not charge and shock.As a result, defibrillation would not be possible if it were necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK CR(R) PLUS DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7732369
Report Number0003015876-2018-01206
Device Sequence Number19229666
Product Code MKJ
UDI-Device Identifier00885074440323
UDI-Public00885074440323
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K033275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCRPLUS
Device Catalogue Number99403-000156
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2017
Initial Report FDA Received Date07/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-