| Model Number |
CRPLUS |
| Medical Device Problem Code |
Failure to Analyze Signal (1539)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/10/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio determined that the cause of the reported issue was process residue on a capacitor, designator c157, located on the analog pcb assembly.This resulted in an improper voltage at integrated circuit (ic) designator u46 on leg 7.The customer was provided with a replacement device.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had a service wrench icon on its readiness display.There was no patient use associated with the reported event.Upon evaluation of the customer¿s device, physio-control found that the device does not recognize a shockable rhythm and does not charge and shock.As a result, defibrillation would not be possible if it were necessary.
|
| |
|