• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE THORACIC; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE THORACIC; PLATE Back to Search Results
Model Number 24-015-22-71
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Reference exemption number e2017029.
 
Event or Problem Description
It was reported two rib plates broke in the patient.
 
Additional Manufacturer Narrative
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The failure root cause cannot be determined due to no device being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.Manufacture date for the lot numbers reported: lots: 33220175 and 33215932; dates 16 oct 2017 and 11 sep 2017.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE THORACIC
Common Device Name
PLATE
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau 78570
GM  78570
MDR Report Key7732788
Report Number9610905-2018-00148
Device Sequence Number1588410
Product Code HRS
UDI-Device Identifier00888118101327
UDI-Public(01)00888118101327
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K153482
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/02/2018
Report Date (Section F) 07/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24-015-22-71
Device Lot Number33220175, 33215932
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date12/05/2018
Event Location Hospital
Date Report to Manufacturer07/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2018
Supplement Date Received by Manufacturer12/05/2018
Initial Report FDA Received Date07/30/2018
Supplement Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-