MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY WALLFLEX ¿ DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY WALLFLEX ¿ DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL Back to Search Results
Model Number M00565010
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.(b)(4).The complainant indicated that the device has been disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex duodenal stent was to be used to treat malignant stricture in the pylorus due to antral gastric cancer during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on (b)(6) 2018.Reportedly, the patient's anatomy was not tortuous and was not dilated prior to stent placement.According to the complainant, during procedure, the stent was attempted to be deployed; however, it was noted that the delivery system did not contain the stent after full deployment.The delivery system was removed from the patient, and another wallflex duodenal stent was placed to complete the procedure.Reportedly, the device was inspected and found no stent on the delivery system; however, a stent was found on the ground.According to the complainant, the stent and delivery system were separated in the package.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX ¿ DUODENAL
Common Device Name
STENT,METALLIC,EXPANDABLE,DUODENAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7732861
Report Number3005099803-2018-02445
Device Sequence Number19173734
Product Code MUM
UDI-Device Identifier08714729456483
UDI-Public08714729456483
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K062750
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/11/2019
Device Model NumberM00565010
Device Catalogue Number6501
Device Lot Number0020630297
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2018
Initial Report FDA Received Date07/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-