| Model Number |
S7 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was unavailable from the site.No parts have been returned to the manufacturer for analysis.Parts not returned.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used for a cranial resection procedure.The issue occurred intraoperatively and delayed the surgery by 20 minutes.It was reported that the em stylet has recognition failure.The em stylet was replaced with a new one and the replaced one could be recognized normally.The surgery was completed with navigation and there was no impact on patient outcome.
|
| |
|