• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COSMAN MEDICAL CSK 10CM TC ELECTRODE WITH 3M CABLE; PROBE, RADIOFREQUENCY LESION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COSMAN MEDICAL CSK 10CM TC ELECTRODE WITH 3M CABLE; PROBE, RADIOFREQUENCY LESION Back to Search Results
Model Number CSK-TC10-3M
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Expiration date: na: health care professional.
 
Event or Problem Description
Device analysis was performed on the returned electrode found that the shaft of the electrode was found to be completely separated from the hub due to excessive external mechanical force.The shaft was not returned for analysis.The functional test could not be performed due to the separated shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSK 10CM TC ELECTRODE WITH 3M CABLE
Common Device Name
PROBE, RADIOFREQUENCY LESION
Manufacturer (Section D)
COSMAN MEDICAL
22 terry ave
burlington MA 01803
MDR Report Key7733140
Report Number3006630150-2018-02289
Device Sequence Number2542252
Product Code GXI
UDI-Device Identifier00813250011524
UDI-Public00813250011524
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K050084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCSK-TC10-3M
Device Lot NumberL490
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2018
Initial Report FDA Received Date07/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-