MEDTRONIC HEART VALVES DIVISION ENVEO R DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number ENVEOR-N-US
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Date of Event 07/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The main component of the system.Other relevant device(s) are: product id: ev olutpro-26-us, serial/lot #: (b)(4), ubd: (b)(6) 2020, udi#: (b)(4) coding for serial number (b)(4), (b)(4) product analysis: the valve remains implanted and the delivery catheter system (dcs) has not yet been received for analysis.Conclusion: without return of the product, no definitive conclusions could be drawn regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that following the implant of this transcatheter bioprosthetic valve, the nose cone of the delivery catheter system (dcs) was not centered and dislodged the valve out of the annulus while the dcs was withdrawn.The valve was snared into the ascending aorta.No additional adverse patient effects were reported for this device.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVEO R DELIVERY SYSTEM
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e. deere ave.
santa ana CA 92705
MDR Report Key7733240
Report Number2025587-2018-01934
Device Sequence Number1018350
Product Code NPT
UDI-Device Identifier00643169800939
UDI-Public00643169800939
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/03/2019
Device Model NumberENVEOR-N-US
Device Catalogue NumberENVEOR-N-US
Device Lot Number0008938616
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/02/2018
Supplement Date Received by Manufacturer08/08/2018
Initial Report FDA Received Date07/30/2018
Supplement Report FDA Received Date10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age89 YR
Patient Weight44
-
-