MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1711K |
Device Problems
No Display/Image (1183); Improper or Incorrect Procedure or Method (2017)
|
Patient Problem
Hyperglycemia (1905)
|
Event Date 06/30/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the paradigm real-time 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported that the insulin pump had a partial display after spilling a drink on the insulin pump.Blood glucose value was 15.0 mmol/l at the time of the incident.The customer went to the hospital due to their high blood glucose value after the insulin pump began having issue so they could receive insulin.The customer provided no additional information regarding their er visit.Troubleshooting was performed but did not resolve the display issue.The device will be replaced.
|
|
Manufacturer Narrative
|
After battery installation device gave an unexpected flashing white or blank display followed by an unexpected intermittent beep alarm due to cracked lcd controller.Unable to perform functional testing including self test, rewind, seating, basic occlusion test, occlusion test, force sensor test and displacement test due to display anomaly.Unable to confirm missing segments/partial display due to display anomaly.Device also received with corroded electronic assembly and motor assembly during visual inspection.
|
|
Manufacturer Narrative
|
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0958-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|